VITROS Chemistry Products Calibrator Kit 29, REF/Product Code 680 2344, used in conjunction with VITROS Chemistry Products dTIBC Reagent Generation 30 (GEN 30), REF 680 2001, and Assay Data Disk (ADD), REF/Product Code 6801876 (Unique Device No. 10758750001576), Data Release Versions (DRVs) 5870 through 5883 inclusive; IVD --- Ortho-Clinical Diagnostics For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 29 is used to calibrate the VITROS 5,1 FS Chemistry System, the VIT

Reported: July 13, 2016 Initiated: May 24, 2016 #Z-2119-2016

Product Description

VITROS Chemistry Products Calibrator Kit 29, REF/Product Code 680 2344, used in conjunction with VITROS Chemistry Products dTIBC Reagent Generation 30 (GEN 30), REF 680 2001, and Assay Data Disk (ADD), REF/Product Code 6801876 (Unique Device No. 10758750001576), Data Release Versions (DRVs) 5870 through 5883 inclusive; IVD --- Ortho-Clinical Diagnostics For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 29 is used to calibrate the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent.

Reason for Recall

Ortho Clinical Diagnostics started receiving customer complaints for biased results when using VITROS dTIBC Reagent GEN 30 product. Ortho's investigation confirmed that incorrect calibration mathematics were assigned to the Calibrator Kit 29 Lots 2915 and 2995 supporting dTIBC Reagent GEN 30 product.

Details

Units Affected
US: 1517; Foreign: 842
Distribution
Worldwide Distribution - US (Nationwide, Puerto Rico) and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, Poland, and Venezuela.
Location
Rochester, NY

Frequently Asked Questions

What product was recalled?
VITROS Chemistry Products Calibrator Kit 29, REF/Product Code 680 2344, used in conjunction with VITROS Chemistry Products dTIBC Reagent Generation 30 (GEN 30), REF 680 2001, and Assay Data Disk (ADD), REF/Product Code 6801876 (Unique Device No. 10758750001576), Data Release Versions (DRVs) 5870 through 5883 inclusive; IVD --- Ortho-Clinical Diagnostics For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 29 is used to calibrate the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent.. Recalled by Ortho-Clinical Diagnostics. Units affected: US: 1517; Foreign: 842.
Why was this product recalled?
Ortho Clinical Diagnostics started receiving customer complaints for biased results when using VITROS dTIBC Reagent GEN 30 product. Ortho's investigation confirmed that incorrect calibration mathematics were assigned to the Calibrator Kit 29 Lots 2915 and 2995 supporting dTIBC Reagent GEN 30 product.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 13, 2016. Severity: Low. Recall number: Z-2119-2016.