PlainRecalls
FDA Devices Moderate Class II Ongoing

Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version: Not Applicable Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. VITROS Performance Verifier is prepared from processed human serum to which enzymes, electrolytes, stabili

Reported: July 23, 2025 Initiated: June 10, 2025 #Z-2120-2025

Product Description

Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version: Not Applicable Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. VITROS Performance Verifier is prepared from processed human serum to which enzymes, electrolytes, stabilizers,mpreservatives and other organic analytes have been added. Component: No, not a component

Reason for Recall

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

Details

Units Affected
37429 units (US)/11,703 units (OUS)
Distribution
Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.
Location
Rochester, NY

Frequently Asked Questions

What product was recalled?
Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version: Not Applicable Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. VITROS Performance Verifier is prepared from processed human serum to which enzymes, electrolytes, stabilizers,mpreservatives and other organic analytes have been added. Component: No, not a component. Recalled by Ortho-Clinical Diagnostics, Inc.. Units affected: 37429 units (US)/11,703 units (OUS).
Why was this product recalled?
Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).
Which agency issued this recall?
This recall was issued by the FDA Devices on July 23, 2025. Severity: Moderate. Recall number: Z-2120-2025.