PlainRecalls
FDA Devices Moderate Class II Terminated

Image1 SPIES X-LINK Camera Control Unit (CCU), Product Model No. TC301, when used with Image1 SPIES D1 camera head, Product Model No. TH116 & TH115, and DC1 Video Mediastinoscope

Reported: July 13, 2016 Initiated: June 20, 2016 #Z-2134-2016

Product Description

Image1 SPIES X-LINK Camera Control Unit (CCU), Product Model No. TC301, when used with Image1 SPIES D1 camera head, Product Model No. TH116 & TH115, and DC1 Video Mediastinoscope

Reason for Recall

Potential electrical interference issue when the Image1 SPIES X-LINK and Image1 SPIES D1 camera head are used in conjunction with the DCI Video Mediastinocope and an electro-surgery generator (ESG).

Details

Recalling Firm
Karl Storz Endoscopy
Units Affected
619 units total (90 units in US)
Distribution
Worldwide Distribution - US Distribution to the states of : IL, NM, TX, MD, TN, PA, FL, KY, MO, NY, MN, DC, MA, WI, CT, MI, ID, CO, GA, NJ and AR., and to the countries of : Argentina, Australia, Belgium, Bulgaria, China, Germany, Finland, Great Britain, India, Italy and Japan.
Location
El Segundo, CA

Frequently Asked Questions

What product was recalled?
Image1 SPIES X-LINK Camera Control Unit (CCU), Product Model No. TC301, when used with Image1 SPIES D1 camera head, Product Model No. TH116 & TH115, and DC1 Video Mediastinoscope. Recalled by Karl Storz Endoscopy. Units affected: 619 units total (90 units in US).
Why was this product recalled?
Potential electrical interference issue when the Image1 SPIES X-LINK and Image1 SPIES D1 camera head are used in conjunction with the DCI Video Mediastinocope and an electro-surgery generator (ESG).
Which agency issued this recall?
This recall was issued by the FDA Devices on July 13, 2016. Severity: Moderate. Recall number: Z-2134-2016.