MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi1000 CONCERTO PIN +FLEXsoft; d. Mil000 CONCERTO +Standard; e. Mil000 CONCERTO +Medium; f. Mi1000 CONCERTO PIN +Standard; g. M11000 CONCERTO +Compressed Product Usage: The Mi1000 CONCERTO Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatibl

Reported: August 14, 2019 Initiated: October 16, 2018 #Z-2136-2019

Product Description

MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi1000 CONCERTO PIN +FLEXsoft; d. Mil000 CONCERTO +Standard; e. Mil000 CONCERTO +Medium; f. Mi1000 CONCERTO PIN +Standard; g. M11000 CONCERTO +Compressed Product Usage: The Mi1000 CONCERTO Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatible MED EL external components. The MED EL Cochlear Implant System is intended to evoke auditory sensations via electrical stimulation of the auditory pathways for severely to profoundly hearing impaired individuals who obtain little or no benefit from acoustic amplification in the best aided condition.

Reason for Recall

Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.

Details

Units Affected
7 devices
Distribution
Worldwide distribution - US Nationwide in the states of California and Illinois. Countries of ARGENTINA, AUSTRALIA, AUSTRIA, BRAZIL, BULGARIA, CHINA, FRANCE, GEORGIA, KOREA, REPUBLIC OF, POLAND, RUSSIAN FEDERATION, SPAIN, SWEDEN.
Location
Innsbruck, N/A

Frequently Asked Questions

What product was recalled?
MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi1000 CONCERTO PIN +FLEXsoft; d. Mil000 CONCERTO +Standard; e. Mil000 CONCERTO +Medium; f. Mi1000 CONCERTO PIN +Standard; g. M11000 CONCERTO +Compressed Product Usage: The Mi1000 CONCERTO Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatible MED EL external components. The MED EL Cochlear Implant System is intended to evoke auditory sensations via electrical stimulation of the auditory pathways for severely to profoundly hearing impaired individuals who obtain little or no benefit from acoustic amplification in the best aided condition.. Recalled by MED-EL Elektromedizinische Gereate, Gmbh. Units affected: 7 devices.
Why was this product recalled?
Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 14, 2019. Severity: Moderate. Recall number: Z-2136-2019.