Infusomat Space Volumetric Pump Administration Set, Product Code 490100

Reported: August 4, 2021 Initiated: June 25, 2021 #Z-2137-2021

Product Description

Infusomat Space Volumetric Pump Administration Set, Product Code 490100

Reason for Recall

There is a potential for the Anti-free flow clip of the administration set to be inverted which may cause unintended free flow if the user fails to follow on screen prompt to close the roller clamp prior to set removal from the pump.

Details

Recalling Firm
B. Braun Medical, Inc.
Units Affected
36,144 units
Distribution
Affected products were distributed to the following US states: CA, MI, OR, and UT.
Location
Allentown, PA

Frequently Asked Questions

What product was recalled?
Infusomat Space Volumetric Pump Administration Set, Product Code 490100. Recalled by B. Braun Medical, Inc.. Units affected: 36,144 units.
Why was this product recalled?
There is a potential for the Anti-free flow clip of the administration set to be inverted which may cause unintended free flow if the user fails to follow on screen prompt to close the roller clamp prior to set removal from the pump.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 4, 2021. Severity: Moderate. Recall number: Z-2137-2021.