PlainRecalls
FDA Devices Moderate Class II Ongoing

Aeos Robotic Digital Microscope, Product Code: PV010

Reported: June 19, 2024 Initiated: May 13, 2024 #Z-2139-2024

Product Description

Aeos Robotic Digital Microscope, Product Code: PV010

Reason for Recall

Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

Details

Recalling Firm
Aesculap Inc
Units Affected
19 units
Distribution
US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.
Location
Center Valley, PA

Frequently Asked Questions

What product was recalled?
Aeos Robotic Digital Microscope, Product Code: PV010. Recalled by Aesculap Inc. Units affected: 19 units.
Why was this product recalled?
Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 19, 2024. Severity: Moderate. Recall number: Z-2139-2024.