The PCA 3 system cam be used in a wide range of clinical settings, including but not limited to: general floor, labor/delivery,post-partum, burn unit, medical surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), and pediatrics. The PCA 3 infusion pump allows clinicians to administer or patients to self-administer analgesia safely and effectively within clinician programmed limits. The epidural route can be used to provide anesthesia or analgesia.
Product Description
The PCA 3 system cam be used in a wide range of clinical settings, including but not limited to: general floor, labor/delivery,post-partum, burn unit, medical surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), and pediatrics. The PCA 3 infusion pump allows clinicians to administer or patients to self-administer analgesia safely and effectively within clinician programmed limits. The epidural route can be used to provide anesthesia or analgesia.
Reason for Recall
Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.
Details
- Recalling Firm
- Hospira Inc.
- Units Affected
- 8,216 infusion pump systems
- Distribution
- Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY) + District of Columbia + Puerto Rico + Virgin Islands; *** FOREIGN:
- Location
- Lake Forest, IL
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11