Hip Products Item Number: 1) 110010462 RingLoc Hip System, Acetabular Bi-Polar Cup, 28 MM X 51 MM 2) 11-165218 RingLoc Bi-Polar Hip System, Acetabular Cup, 28 MM X 47 MM 3) XL-105916 588500 RingLoc Hip System, Acetabular Liner, 36 MM, Size 26 Product Usage: Hemi hip arthroplasty.

Reported: June 3, 2020 Initiated: April 17, 2020 #Z-2145-2020

Product Description

Hip Products Item Number: 1) 110010462 RingLoc Hip System, Acetabular Bi-Polar Cup, 28 MM X 51 MM 2) 11-165218 RingLoc Bi-Polar Hip System, Acetabular Cup, 28 MM X 47 MM 3) XL-105916 588500 RingLoc Hip System, Acetabular Liner, 36 MM, Size 26 Product Usage: Hemi hip arthroplasty.

Reason for Recall

Potential presence of elevated endotoxin levels that exceed the specification limit

Details

Recalling Firm
Biomet, Inc.
Units Affected
58 units
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, AHMADABAD, GUJARAT, BANGKOK, BELROSE NSW, BUENOS AIRES, CHILE, COSTA RICA, HIRATSUKA CITY, HUECHURABA SANTIAGO, INDIA, JAPAN, MEXICO DF, NETHERLANDS, PICHINCHA, EC, SAN JOSE, SINGAPORE, TAIWAN, THAILAND, YONGSAN-GU SEOUL.
Location
Warsaw, IN

Frequently Asked Questions

What product was recalled?
Hip Products Item Number: 1) 110010462 RingLoc Hip System, Acetabular Bi-Polar Cup, 28 MM X 51 MM 2) 11-165218 RingLoc Bi-Polar Hip System, Acetabular Cup, 28 MM X 47 MM 3) XL-105916 588500 RingLoc Hip System, Acetabular Liner, 36 MM, Size 26 Product Usage: Hemi hip arthroplasty.. Recalled by Biomet, Inc.. Units affected: 58 units.
Why was this product recalled?
Potential presence of elevated endotoxin levels that exceed the specification limit
Which agency issued this recall?
This recall was issued by the FDA Devices on June 3, 2020. Severity: Moderate. Recall number: Z-2145-2020.