PlainRecalls
FDA Devices Moderate Class II Terminated

Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100

Reported: August 14, 2019 Initiated: June 18, 2019 #Z-2160-2019

Product Description

Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100

Reason for Recall

Discrepant length; Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length.

Details

Recalling Firm
Howmedica Osteonics Corp.
Units Affected
11
Distribution
US Nationwide Distribution - CA, FL, GA, MA, NY, PA, TX
Location
Allendale, NJ

Frequently Asked Questions

What product was recalled?
Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100. Recalled by Howmedica Osteonics Corp.. Units affected: 11.
Why was this product recalled?
Discrepant length; Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 14, 2019. Severity: Moderate. Recall number: Z-2160-2019.