FlexLab Augomation System using the Aliquoter Module (Inpeco P/N FLX-212) The systems consolidate multiple Analytical instruments into a unified workstation.

Reported: June 10, 2020 Initiated: April 7, 2020 #Z-2198-2020

Product Description

FlexLab Augomation System using the Aliquoter Module (Inpeco P/N FLX-212) The systems consolidate multiple Analytical instruments into a unified workstation.

Reason for Recall

Potential for discrepant patient results due to a potential for the primary sample tube to be diluted. In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.

Details

Recalling Firm
Inpeco S.A.
Units Affected
71 systems
Distribution
US Nationwide distribution including in the states of IL, NY.
Location
Lugano, N/A

Frequently Asked Questions

What product was recalled?
FlexLab Augomation System using the Aliquoter Module (Inpeco P/N FLX-212) The systems consolidate multiple Analytical instruments into a unified workstation.. Recalled by Inpeco S.A.. Units affected: 71 systems.
Why was this product recalled?
Potential for discrepant patient results due to a potential for the primary sample tube to be diluted. In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 10, 2020. Severity: Moderate. Recall number: Z-2198-2020.