AEQUALIS Reversed II or AEQUALIS Reversed Fracture Impactor. These products are contained within the Aequalis reversed II Instrument set tray no.YKAD83 and the Aequalis reversed fracture instrument set tray no.YKAD95. The items affected are MWD21, MWD023, MWD024 and MWD025. The impactor is an instrument used during shoulder arthroplasty; it is not implanted. The impactor is used to set the glenoid sphere, poly insert, spacer, and hemi adapter into place during surgery. The impactor is a t

Reported: September 25, 2013 Initiated: August 13, 2013 #Z-2199-2013

Product Description

AEQUALIS Reversed II or AEQUALIS Reversed Fracture Impactor. These products are contained within the Aequalis reversed II Instrument set tray no.YKAD83 and the Aequalis reversed fracture instrument set tray no.YKAD95. The items affected are MWD21, MWD023, MWD024 and MWD025. The impactor is an instrument used during shoulder arthroplasty; it is not implanted. The impactor is used to set the glenoid sphere, poly insert, spacer, and hemi adapter into place during surgery. The impactor is a two piece instrument which consists of a tip and a handle. The tip is made of plastic and metal that is pressed together for assembly.

Reason for Recall

Following the identification of the disassociation of an impactor tip during surgery, Tornier is initiating a voluntary recall of all lots of Aequalis Reversed II and Reversed Fracture impactors. Upon review of the event, Tornier determined that a redesign of the instrument was warranted. These products are contained within the Aequalis Revered II instrument set tray no YKAD83 and the Aequalis R

Details

Recalling Firm
Tornier, Inc
Units Affected
2,298 (1045 US, 1253 OUS)
Distribution
.Worldwide Distribution - USA including Puerto Rico and the states of AL, AZ, CA, CO,CT, FL, GA, IL, KY, MD, MA, MI, MN, NE, NJ, NC, OH, OR, PA and TX., and the countries of CANADA, MEXICO, HONG KONG, SOUTH AFRICA, ARGENTINA, COLOMBIA, ISRAEL, AUSTRALIA, BELGIUM, FINLAND,FRANCE, GERMANY, ITALY, NETHERLANDS, SLOVENIA, SPAIN, SWITZERLAND and UNITED KINGDOM
Location
Bloomington, MN

Frequently Asked Questions

What product was recalled?
AEQUALIS Reversed II or AEQUALIS Reversed Fracture Impactor. These products are contained within the Aequalis reversed II Instrument set tray no.YKAD83 and the Aequalis reversed fracture instrument set tray no.YKAD95. The items affected are MWD21, MWD023, MWD024 and MWD025. The impactor is an instrument used during shoulder arthroplasty; it is not implanted. The impactor is used to set the glenoid sphere, poly insert, spacer, and hemi adapter into place during surgery. The impactor is a two piece instrument which consists of a tip and a handle. The tip is made of plastic and metal that is pressed together for assembly.. Recalled by Tornier, Inc. Units affected: 2,298 (1045 US, 1253 OUS).
Why was this product recalled?
Following the identification of the disassociation of an impactor tip during surgery, Tornier is initiating a voluntary recall of all lots of Aequalis Reversed II and Reversed Fracture impactors. Upon review of the event, Tornier determined that a redesign of the instrument was warranted. These products are contained within the Aequalis Revered II instrument set tray no YKAD83 and the Aequalis R
Which agency issued this recall?
This recall was issued by the FDA Devices on September 25, 2013. Severity: Moderate. Recall number: Z-2199-2013.