Philips Azurion systems with software version R1.2 -Interventional Fluoroscopic X-ray system as follows: Azurion 3 M12, Azurion 3 M15, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20 Models: 722063, 722064, 722067, 722068, 722078, 722079. Product Usage: The Azurion series (within the limits of the used Operating Room table) are intended for use to perform: Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical app
Product Description
Philips Azurion systems with software version R1.2 -Interventional Fluoroscopic X-ray system as follows: Azurion 3 M12, Azurion 3 M15, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20 Models: 722063, 722064, 722067, 722068, 722078, 722079. Product Usage: The Azurion series (within the limits of the used Operating Room table) are intended for use to perform: Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures. Additionally: . The Azurion series can be used in a hybrid Operating Room. The Azurion series contain a number of features to support a flexible and patient centric procedural workflow.
Reason for Recall
The cold restart of Azurion R1.2 systems may take up to 7 minutes if the system is connected to the mains power supply for more than 50 days, may result in a delay of treatment.
Details
- Recalling Firm
- Philips North America, LLC
- Units Affected
- 834
- Distribution
- US Nationwide Distribution Foreign: Canada Argentina AUSTRALIA Austria Azerbaijan Bahrain Bangladesh Belgium Bosnia and Herzegovina Brazil Bulgaria Chile Colombia CROATIA Denmark ECUADOR Egypt Finland France Germany Greece Hong Kong India Indonesia Ireland Israel Italie Italy Japan Jordan Kazakhstan Korea, Republic of Latvia Lebanon Libya Malaysia Maldives Mexico Mongolia Myanmar Netherlands New Zealand Norway Oman Pakistan Philippines Polamnd Poland Portugal Republic of Korea Romania Russian Federation Saudi Arabia Serbia Singapore SLOVENIA South Africa Spain Sri Lanka Sweden Switzerland Thailand Tunisia Turkey UNITED ARAB EMI United Arab Emirates United Kingdom Uzbekistan Viet Nam
- Location
- Andover, MA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11