FDA Devices Moderate Class II Ongoing

Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 500e Blood Gas System (China) 11416752; RAPIDPoint 500e Blood Gas System (Japan) 11416754; RAPIDPoint 500e Blood Gas System (ROW) 11416755 ;

Reported: July 3, 2024 Initiated: May 17, 2024 #Z-2214-2024

Product Description

Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 500e Blood Gas System (China) 11416752; RAPIDPoint 500e Blood Gas System (Japan) 11416754; RAPIDPoint 500e Blood Gas System (ROW) 11416755 ;

Reason for Recall

Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software version, samples run in the capillary mode are incorrectly labeled as arterial when displayed on the LIS, leading to the potential for results to be flagged according to an established arterial range. Capillary sample results are correctly identified as capillary on both the RAPIDPoint 500e system display and the instrument printouts. The issue only affects the new software version 5.3 and only capillary sample mode is impacted. The inability to distinguish specimen sources for blood gas may lead to incorrect diagnoses and management decisions.

Details

Units Affected
9 units US
Distribution
Worldwide - US Nationwide distribution in the states of FL, GA, LA, MI, NC and the countries of Chile, Costa Rica, France, Germany, Japan, Lebanon, Mexico, Netherlands, New Zealand, Poland, Spain, Turkey, United Kingdom.
Location
Norwood, MA

Frequently Asked Questions

What product was recalled?
Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 500e Blood Gas System (China) 11416752; RAPIDPoint 500e Blood Gas System (Japan) 11416754; RAPIDPoint 500e Blood Gas System (ROW) 11416755 ;. Recalled by Siemens Healthcare Diagnostics Inc. Units affected: 9 units US.
Why was this product recalled?
Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software version, samples run in the capillary mode are incorrectly labeled as arterial when displayed on the LIS, leading to the potential for results to be flagged according to an established arterial range. Capillary sample results are correctly identified as capillary on both the RAPIDPoint 500e system display and the instrument printouts. The issue only affects the new software version 5.3 and only capillary sample mode is impacted. The inability to distinguish specimen sources for blood gas may lead to incorrect diagnoses and management decisions.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 3, 2024. Severity: Moderate. Recall number: Z-2214-2024.