TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES D684-00-0545-01 TRANS-RAY 7FR. 40cc IAB WITH ACCESSORIES D684-00-0546-01 TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES D684-00-0607 TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES D684-00-0608 PACKAGED INSERTION KIT - TRANS-RAY 7 Fr. 34/40 cc IABs D

Reported: August 2, 2023 Initiated: June 23, 2023 #Z-2218-2023

Product Description

TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES D684-00-0545-01 TRANS-RAY 7FR. 40cc IAB WITH ACCESSORIES D684-00-0546-01 TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES D684-00-0607 TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES D684-00-0608 PACKAGED INSERTION KIT - TRANS-RAY 7 Fr. 34/40 cc IABs D884-00-0019-18 PACKAGED INSERTION KIT - TRANS-RAY PLUS 7.5 Fr. 35/40 IABs D884-00-0019-26

Reason for Recall

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Details

Recalling Firm
Datascope Corp.
Units Affected
321,609 total kits
Distribution
Domestic distribution nationwide. International distribution worldwide.
Location
Fairfield, NJ

Frequently Asked Questions

What product was recalled?
TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES D684-00-0545-01 TRANS-RAY 7FR. 40cc IAB WITH ACCESSORIES D684-00-0546-01 TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES D684-00-0607 TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES D684-00-0608 PACKAGED INSERTION KIT - TRANS-RAY 7 Fr. 34/40 cc IABs D884-00-0019-18 PACKAGED INSERTION KIT - TRANS-RAY PLUS 7.5 Fr. 35/40 IABs D884-00-0019-26. Recalled by Datascope Corp.. Units affected: 321,609 total kits.
Why was this product recalled?
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 2, 2023. Severity: Moderate. Recall number: Z-2218-2023.