PlainRecalls
FDA Devices Moderate Class II Ongoing

SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES D684-00-0568-01 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (W/O STATLOCK) D684-00-0568-03 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (WITH PRESSURE D684-00-0568-05 SENSATION PLUS 7.5Fr. 40cc WITH

Reported: August 2, 2023 Initiated: June 23, 2023 #Z-2221-2023

Product Description

SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES D684-00-0568-01 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (W/O STATLOCK) D684-00-0568-03 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (WITH PRESSURE D684-00-0568-05 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (TURKEY) D684-00-0568-09 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES, US ONLY D684-00-0568-01U PACKAGED INSERTION KIT - SENSATION PLUS 7.5 Fr. 40 cc IAB D884-00-0019-22

Reason for Recall

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Details

Recalling Firm
Datascope Corp.
Units Affected
321,609 total kits
Distribution
Domestic distribution nationwide. International distribution worldwide.
Location
Fairfield, NJ

Frequently Asked Questions

What product was recalled?
SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES D684-00-0568-01 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (W/O STATLOCK) D684-00-0568-03 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (WITH PRESSURE D684-00-0568-05 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (TURKEY) D684-00-0568-09 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES, US ONLY D684-00-0568-01U PACKAGED INSERTION KIT - SENSATION PLUS 7.5 Fr. 40 cc IAB D884-00-0019-22. Recalled by Datascope Corp.. Units affected: 321,609 total kits.
Why was this product recalled?
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 2, 2023. Severity: Moderate. Recall number: Z-2221-2023.