Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.

Reported: August 6, 2025 Initiated: June 13, 2025 #Z-2221-2025

Product Description

Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.

Reason for Recall

Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire"

Details

Recalling Firm
Volcano Corp
Units Affected
233,817
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of BR, PL, SK, IT, DK, DE, GB, AT, KW, CR, ES, NL, LV, MT, IE, CH, ZA, RO, FI, PR, FR, CA, PA, IL, SE, PT, HR, GR, NZ, HK, TW, MY, TH, MX, AR, PE, AE, BG, AU, JP, ID, RS, CO, SG, BE.
Location
San Diego, CA

Frequently Asked Questions

What product was recalled?
Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.. Recalled by Volcano Corp. Units affected: 233,817.
Why was this product recalled?
Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire"
Which agency issued this recall?
This recall was issued by the FDA Devices on August 6, 2025. Severity: Moderate. Recall number: Z-2221-2025.