PlainRecalls
FDA Devices Moderate Class II Ongoing

Flexiva Pulse ID-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy (Box of 5)UPN: M006L8406911

Reported: August 2, 2023 Initiated: May 25, 2023 #Z-2232-2023

Product Description

Flexiva Pulse ID-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy (Box of 5)UPN: M006L8406911

Reason for Recall

Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched

Details

Units Affected
205 units
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Location
Marlborough, MA

Frequently Asked Questions

What product was recalled?
Flexiva Pulse ID-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy (Box of 5)UPN: M006L8406911. Recalled by Boston Scientific Corporation. Units affected: 205 units.
Why was this product recalled?
Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched
Which agency issued this recall?
This recall was issued by the FDA Devices on August 2, 2023. Severity: Moderate. Recall number: Z-2232-2023.