Computed Tomography X-ray Systems Ingenuity Core. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Reported: August 5, 2015 Initiated: February 5, 2015 #Z-2257-2015

Product Description

Computed Tomography X-ray Systems Ingenuity Core. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Reason for Recall

5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value incorrectly remains constant for some heart rate change conditions. DoseRight incorrectly enabled when the Exam Card specified disabling DoseRight. Tube heat predict

Details

Units Affected
82 units
Distribution
Worldwide Distribution-US (nationwide) including the states of AR, AZ, CA, CT, DE, FL, GA, IA, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VT, WV & WY, and to the following countries: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cambodia, Canada, China, Colombia, Denmark, Finland, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Italy, Japan, Latvia, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Norway, Panama, Philippines, Russia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan & Viet Nam.
Location
Cleveland, OH

Frequently Asked Questions

What product was recalled?
Computed Tomography X-ray Systems Ingenuity Core. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.. Recalled by Philips Medical Systems (Cleveland) Inc. Units affected: 82 units.
Why was this product recalled?
5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value incorrectly remains constant for some heart rate change conditions. DoseRight incorrectly enabled when the Exam Card specified disabling DoseRight. Tube heat predict
Which agency issued this recall?
This recall was issued by the FDA Devices on August 5, 2015. Severity: Moderate. Recall number: Z-2257-2015.