Sheridan/HVT Novaplus: a) 7.0 mm, REF V5-10314 b) 7.5 mm, REF V5-10315 c) 8.0 mm, REF V5-10316 d) 8.5 mm, REF V5-10317 Product Usage: Tracheal tube/airway management

Reported: September 18, 2019 Initiated: July 30, 2019 #Z-2318-2019

Product Description

Sheridan/HVT Novaplus: a) 7.0 mm, REF V5-10314 b) 7.5 mm, REF V5-10315 c) 8.0 mm, REF V5-10316 d) 8.5 mm, REF V5-10317 Product Usage: Tracheal tube/airway management

Reason for Recall

Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.

Details

Recalling Firm
Teleflex Medical
Units Affected
1391004 units
Distribution
US Nationwide Distribution
Location
Morrisville, NC

Frequently Asked Questions

What product was recalled?
Sheridan/HVT Novaplus: a) 7.0 mm, REF V5-10314 b) 7.5 mm, REF V5-10315 c) 8.0 mm, REF V5-10316 d) 8.5 mm, REF V5-10317 Product Usage: Tracheal tube/airway management. Recalled by Teleflex Medical. Units affected: 1391004 units.
Why was this product recalled?
Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 18, 2019. Severity: Critical. Recall number: Z-2318-2019.