QIAsymphony SP SOW 5.0.3, software used with the QIAsymphony SP Instrument (The QIAsymphony SP instrument is designed to perform automated purification of nucleic acids-IVD) Model Number: 9001297 Product Usage: The QIAsymphony SP instrument is designed to perform automated purification of nucleic acids. It is intended to be used only in combination with QIAsymphony kits indicated for use with the QIAsymphony SP for the applications described in the kit handbooks. The QIAsymphony SP is inten
Product Description
QIAsymphony SP SOW 5.0.3, software used with the QIAsymphony SP Instrument (The QIAsymphony SP instrument is designed to perform automated purification of nucleic acids-IVD) Model Number: 9001297 Product Usage: The QIAsymphony SP instrument is designed to perform automated purification of nucleic acids. It is intended to be used only in combination with QIAsymphony kits indicated for use with the QIAsymphony SP for the applications described in the kit handbooks. The QIAsymphony SP is intended for use by professional users, such as technicians and physicians trained in molecular biological techniques and the operation of the QIAsymphony SP instrument
Reason for Recall
QIAsymphony software version 5.0.3 - Software issue incorreclty assigns the 2D bar code eluate ID to the sample within the results file. If 2D bar code eluate IDs are not checked against the sample ID prior to downstream application the incorrect tube may be selected and it could have the potential to lead to delayed results or incorrect patient reporting
Details
- Recalling Firm
- Qiagen Sciences LLC
- Units Affected
- 417
- Distribution
- Worldwide Distribution - US Nationwide Foreign: Algeria Australia Austria Azerbaijan Belgium Brazil Canada China Czech Republic Denmark Finland France Germany Greece India Israel Italy Japan Jordan Korea, Republic of Kuwait Latvia Mexico Netherlands Norway Pakistan Poland Republic of Korea Russian Federation Singapore South Africa Spain Sweden Switzerland Thailand Turkey United Kingdom
- Location
- Germantown, MD
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11