V-TEK TWIST D 40X90MM 30MM Cann 12MM Rnd; The twist drill is used to drill holes into bones/tissues for foot, ankle and hand implant systems.

Reported: August 10, 2016 Initiated: June 27, 2016 #Z-2324-2016

Product Description

V-TEK TWIST D 40X90MM 30MM Cann 12MM Rnd; The twist drill is used to drill holes into bones/tissues for foot, ankle and hand implant systems.

Reason for Recall

Zimmer Biomet initiated a voluntary recall of the Twist Drill & V-Tek Twist Drill due to to a potential tolerance stack up identified between the twist drill and the mating Kirschner Wire (K-Wire) component.

Details

Recalling Firm
Zimmer Gmbh
Units Affected
169
Distribution
Domestic: CO, IL Foreign: Argentina, Costa Rica, India, Malta, Japan, Philippines, Vietnam, Portugal, Dominican Republic, Argentina, Australia, Russia, Saudi Arabia, Ireland, Ecuador, China, Singapore, Indonesia, Canada, Mexico, South Korea , Spain, Netherlands, Sweden, Malaysia, New Zealand, Switzerland, El Salvador, England, Norway, Finland, Venezuela, Taiwan, Italy, Colombia, Greece, Thailand, France, Israel, Jamaica, Japan, Jordan, Germany, Korea, Lebanon, Belgium, Hong Kong, Philippines, Iran, Brazil, Turkey, United Arab Emirates (UAE), Uruguay
Location
Winterthur

Frequently Asked Questions

What product was recalled?
V-TEK TWIST D 40X90MM 30MM Cann 12MM Rnd; The twist drill is used to drill holes into bones/tissues for foot, ankle and hand implant systems.. Recalled by Zimmer Gmbh. Units affected: 169.
Why was this product recalled?
Zimmer Biomet initiated a voluntary recall of the Twist Drill & V-Tek Twist Drill due to to a potential tolerance stack up identified between the twist drill and the mating Kirschner Wire (K-Wire) component.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 10, 2016. Severity: Moderate. Recall number: Z-2324-2016.