Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP). Model Number: WA91327U. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.

Reported: August 27, 2025 Initiated: July 25, 2025 #Z-2338-2025

Product Description

Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP). Model Number: WA91327U. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.

Reason for Recall

Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.

Details

Units Affected
155 units
Distribution
US Nationwide distribution.
Location
Center Valley, PA

Frequently Asked Questions

What product was recalled?
Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP). Model Number: WA91327U. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.. Recalled by Olympus Corporation of the Americas. Units affected: 155 units.
Why was this product recalled?
Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 27, 2025. Severity: Moderate. Recall number: Z-2338-2025.