LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 25 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-25/00
Reported: August 16, 2023 Initiated: July 3, 2023 #Z-2356-2023
Product Description
LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 25 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-25/00
Reason for Recall
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
Details
- Recalling Firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Units Affected
- 24 units
- Distribution
- US Nationwide distribution.
- Location
- Norderstedt, N/A
Frequently Asked Questions
What product was recalled? ▼
LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 25 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-25/00. Recalled by Waldemar Link GmbH & Co. KG (Mfg Site). Units affected: 24 units.
Why was this product recalled? ▼
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on August 16, 2023. Severity: Moderate. Recall number: Z-2356-2023.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11