PlainRecalls
FDA Devices Moderate Class II Ongoing

Ingenia Elition X Magnetic Resonance (MR) system, Product Numbers 781358, 782107, and 782118 (SmartPath to Ingenia Elition X)

Reported: August 16, 2023 Initiated: June 12, 2023 #Z-2395-2023

Product Description

Ingenia Elition X Magnetic Resonance (MR) system, Product Numbers 781358, 782107, and 782118 (SmartPath to Ingenia Elition X)

Reason for Recall

The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.

Details

Recalling Firm
Philips North America
Units Affected
640 total
Distribution
US Nationwide. Global Distribution.
Location
Cambridge, MA

Frequently Asked Questions

What product was recalled?
Ingenia Elition X Magnetic Resonance (MR) system, Product Numbers 781358, 782107, and 782118 (SmartPath to Ingenia Elition X). Recalled by Philips North America. Units affected: 640 total.
Why was this product recalled?
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
Which agency issued this recall?