PlainRecalls
FDA Devices Moderate Class II Terminated

Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.

Reported: August 17, 2016 Initiated: May 11, 2016 #Z-2404-2016

Product Description

Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.

Reason for Recall

There is 5 mL printed next to the increment on the medicine cups instead of 10 mL. The cup has the correct mL markings for the other gradations, but sequentially, the markings are listed as 2.5 mL, 5 mL, 7.5 mL, and 5 mL.

Details

Recalling Firm
Medline Industries Inc
Units Affected
903 Cases and 4,090 Sleeves (8,605,000 Medicine Cups)
Distribution
Continental U.S.
Location
Mundelein, IL

Frequently Asked Questions

What product was recalled?
Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.. Recalled by Medline Industries Inc. Units affected: 903 Cases and 4,090 Sleeves (8,605,000 Medicine Cups).
Why was this product recalled?
There is 5 mL printed next to the increment on the medicine cups instead of 10 mL. The cup has the correct mL markings for the other gradations, but sequentially, the markings are listed as 2.5 mL, 5 mL, 7.5 mL, and 5 mL.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 17, 2016. Severity: Moderate. Recall number: Z-2404-2016.