Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.

Reported: September 3, 2025 Initiated: July 14, 2025 #Z-2420-2025

Product Description

Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.

Reason for Recall

Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.

Details

Recalling Firm
MAQUET GMBH
Units Affected
315 units (8 US, 307 OUS)
Distribution
Worldwide - US Nationwide distribution in the states of KY, NY, TX and the countries of Algeria, Australia, Austria, Bulgaria, Canada, China, Colombia, Denmark, Egypt, Estonia, Finland, France, Germany, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of, Kuwait, Lebanon, Morocco, Netherlands, Norway, Poland, Portugal, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom.
Location
Rastatt

Frequently Asked Questions

What product was recalled?
Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.. Recalled by MAQUET GMBH. Units affected: 315 units (8 US, 307 OUS).
Why was this product recalled?
Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 3, 2025. Severity: Moderate. Recall number: Z-2420-2025.