WATCHMAN FXD Curve Access Sys Sgl, OUS, Material Number (UPN) M635TS80010; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.

Reported: September 10, 2025 Initiated: July 29, 2025 #Z-2429-2025

Product Description

WATCHMAN FXD Curve Access Sys Sgl, OUS, Material Number (UPN) M635TS80010; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.

Reason for Recall

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

Details

Units Affected
340185 units in total
Distribution
Worldwide.
Location
Maple Grove, MN

Frequently Asked Questions

What product was recalled?
WATCHMAN FXD Curve Access Sys Sgl, OUS, Material Number (UPN) M635TS80010; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.. Recalled by Boston Scientific Corporation. Units affected: 340185 units in total.
Why was this product recalled?
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 10, 2025. Severity: Critical. Recall number: Z-2429-2025.