RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778
Reported: July 1, 2020 Initiated: May 27, 2020 #Z-2432-2020
Product Description
RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778
Reason for Recall
The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.
Details
- Recalling Firm
- Siemens Healthcare Diagnostics Inc
- Units Affected
- 436
- Distribution
- Foreign distribution only.
- Location
- Norwood, MA
Frequently Asked Questions
What product was recalled? ▼
RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778. Recalled by Siemens Healthcare Diagnostics Inc. Units affected: 436.
Why was this product recalled? ▼
The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on July 1, 2020. Severity: Moderate. Recall number: Z-2432-2020.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11