BD Vacutainer SST Plus Blood Collection Tubes, 5.0 mL, 13 x 100 mm, Sterile. Provides a means of collecting, transporting, separating, and processing blood in a closed tube.

Reported: August 26, 2015 Initiated: June 23, 2015 #Z-2434-2015

Product Description

BD Vacutainer SST Plus Blood Collection Tubes, 5.0 mL, 13 x 100 mm, Sterile. Provides a means of collecting, transporting, separating, and processing blood in a closed tube.

Reason for Recall

Some of the tubes were manufactured with stoppers that did not meet current manufacturing specifications. Stoppers were manufactured with a reduced cycle cure time.

Details

Units Affected
238, 000 (Lot #5133959) and 243,000 (Lot #5133957)
Distribution
Distributed in FL, IL, MA, MI, MO, PA and TN.
Location
Franklin Lakes, NJ