Dimension Vista B¿2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of B¿2-Microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista System. Measurements of B¿2-Microglobulin aid in the diagnosis of active rheumatoid arthritis and kidney disease.

Reported: September 2, 2015 Initiated: July 23, 2015 #Z-2450-2015

Product Description

Dimension Vista B¿2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of B¿2-Microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista System. Measurements of B¿2-Microglobulin aid in the diagnosis of active rheumatoid arthritis and kidney disease.

Reason for Recall

customers are receiving a high number of Abnormal Assay Errors or are not able to calibrate B2MIC when using Flex reagent cartridge B2MIC various lots. In failure mode, there is the potential for falsely depressed beta2-Microglobulin results on the Dimension Vista system due to assay reagent contamination.

Details

Units Affected
Total 1655 units
Distribution
US Nationwide Distribution in the states of TN, AZ, CT, Fl, CA, NY and WI
Location
Newark, DE

Frequently Asked Questions

What product was recalled?
Dimension Vista B¿2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of B¿2-Microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista System. Measurements of B¿2-Microglobulin aid in the diagnosis of active rheumatoid arthritis and kidney disease.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: Total 1655 units.
Why was this product recalled?
customers are receiving a high number of Abnormal Assay Errors or are not able to calibrate B2MIC when using Flex reagent cartridge B2MIC various lots. In failure mode, there is the potential for falsely depressed beta2-Microglobulin results on the Dimension Vista system due to assay reagent contamination.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 2, 2015. Severity: Moderate. Recall number: Z-2450-2015.