Hologic Fluoroscan Mini C-Arm Model with InSight2, Software version 4.0 and 4.0.1 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.

Reported: October 3, 2012 Initiated: September 14, 2012 #Z-2452-2012

Product Description

Hologic Fluoroscan Mini C-Arm Model with InSight2, Software version 4.0 and 4.0.1 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.

Reason for Recall

InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180¿around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180¿ around vertical axis as well as 180¿ around a horizontal axis.

Details

Recalling Firm
Hologic, Inc.
Units Affected
414 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia,Belgium, Canada, China, Germany, Ecuador, Spain, France, Great Britain, Greece, Hong Kong, Italy, Japan Korea, Lithuvania, Neherlands, qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan, South Africa, and Switzerland.
Location
Bedford, MA

Frequently Asked Questions

What product was recalled?
Hologic Fluoroscan Mini C-Arm Model with InSight2, Software version 4.0 and 4.0.1 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.. Recalled by Hologic, Inc.. Units affected: 414 units.
Why was this product recalled?
InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180¿around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180¿ around vertical axis as well as 180¿ around a horizontal axis.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 3, 2012. Severity: Moderate. Recall number: Z-2452-2012.