Micro Joystick R-net. Electrical wheelchair component.
Reported: September 10, 2025 Initiated: July 10, 2025 #Z-2454-2025
Product Description
Micro Joystick R-net. Electrical wheelchair component.
Reason for Recall
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Details
- Recalling Firm
- mo-Vis BVBA
- Units Affected
- 3 units
- Distribution
- US Nationwide distribution in the state of TX.
- Location
- Nevele, N/A
Frequently Asked Questions
What product was recalled? ▼
Micro Joystick R-net. Electrical wheelchair component.. Recalled by mo-Vis BVBA. Units affected: 3 units.
Why was this product recalled? ▼
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on September 10, 2025. Severity: Critical. Recall number: Z-2454-2025.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11