PlainRecalls
FDA Devices Moderate Class II Ongoing

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) BORN ON ARRIVAL KIT, Model/Item Number LLBN11C; 2) LABOR & DELIVERY PACK, Model/Item Number LLLD19I.

Reported: September 3, 2025 Initiated: July 14, 2025 #Z-2469-2025

Product Description

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) BORN ON ARRIVAL KIT, Model/Item Number LLBN11C; 2) LABOR & DELIVERY PACK, Model/Item Number LLLD19I.

Reason for Recall

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Details

Units Affected
781 units
Distribution
US distribution to states of: IL and LA
Location
Tiffin, OH

Frequently Asked Questions

What product was recalled?
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) BORN ON ARRIVAL KIT, Model/Item Number LLBN11C; 2) LABOR & DELIVERY PACK, Model/Item Number LLLD19I.. Recalled by American Contract Systems Inc. Units affected: 781 units.
Why was this product recalled?
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 3, 2025. Severity: Moderate. Recall number: Z-2469-2025.