Atellica¿ CH HDL Cholesterol- In vitro diagnostic use in the quantitative determination of HDL cholesterol in human serum and plasma (lithium heparin, EDTA, sodium Heparin) Test Code: HDLC Siemens Material Number (SMN): 11537213
Product Description
Atellica¿ CH HDL Cholesterol- In vitro diagnostic use in the quantitative determination of HDL cholesterol in human serum and plasma (lithium heparin, EDTA, sodium Heparin) Test Code: HDLC Siemens Material Number (SMN): 11537213
Reason for Recall
Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica¿ CH Total Bilirubin_2 (TBil_2), Atellica¿ CH LDL Cholesterol (LDLC), Atellica¿ CH Gamma-Glutamyl Transferase (GGT) and Atellica CH HDL Cholesterol (HDLC)
Details
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc.
- Units Affected
- 18524 units
- Distribution
- US Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay Vietnam
- Location
- Tarrytown, NY
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11