epoc BGEM Crea Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1

Reported: September 6, 2023 Initiated: June 1, 2023 #Z-2477-2023

Product Description

epoc BGEM Crea Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1

Reason for Recall

There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.

Details

Units Affected
473 ea
Distribution
US distribution to AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, KS, KY, LA, MI, MN, MO, NC, ND, NH, NY, OH, OK, SC, SD, and TX OUS distribution to Canada, China, Colombia, Finland, India, Mexico, Norway, Sweden, United Arab Emirates, and Vietnam
Location
Norwood, MA

Frequently Asked Questions

What product was recalled?
epoc BGEM Crea Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1. Recalled by Siemens Healthcare Diagnostics Inc. Units affected: 473 ea.
Why was this product recalled?
There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 6, 2023. Severity: Moderate. Recall number: Z-2477-2023.