PlainRecalls
FDA Devices Moderate Class II Ongoing

Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300

Reported: August 14, 2024 Initiated: June 28, 2024 #Z-2512-2024

Product Description

Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300

Reason for Recall

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport.

Details

Units Affected
400 units
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Location
Deerfield, IL

Frequently Asked Questions

What product was recalled?
Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300. Recalled by Baxter Healthcare Corporation. Units affected: 400 units.
Why was this product recalled?
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 14, 2024. Severity: Moderate. Recall number: Z-2512-2024.