NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662

Reported: September 10, 2025 Initiated: June 25, 2025 #Z-2514-2025

Product Description

NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662

Reason for Recall

Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation, resulting in restoration delay.

Details

Recalling Firm
Preat Corp
Units Affected
11
Distribution
US Nationwide distribution in the states of NY, FL, PA, CA, OH.
Location
Santa Maria, CA

Frequently Asked Questions

What product was recalled?
NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662. Recalled by Preat Corp. Units affected: 11.
Why was this product recalled?
Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation, resulting in restoration delay.