Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 Product Description: The Merge Hemo, Model RCSV2 device is a hemodynamics recording computer system that monitors, measures, displays, records and stores various physiologic and blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices. The system is used for documenting and monitoring patients undergoing diagnostic and interventional catheter-based procedu
Product Description
Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 Product Description: The Merge Hemo, Model RCSV2 device is a hemodynamics recording computer system that monitors, measures, displays, records and stores various physiologic and blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices. The system is used for documenting and monitoring patients undergoing diagnostic and interventional catheter-based procedures. The procedures can be performed in the clinical areas of invasive cardiology, interventional radiology, and cardiac electrophysiology.
Reason for Recall
Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client (Pressure Sensor Defective) during non-invasive blood pressure (NIBP) measurements, specifically when the initial inflation pressure is set in the high-pressure range (>240 mmHg). The ARGUS PB-3000 is a vital signs monitoring instrument intended to be integrated into the Merge Hemo System. No other components of the Merge Hemo System are affected.
Details
- Recalling Firm
- Merge Healthcare, Inc.
- Units Affected
- 84
- Distribution
- US Nationwide distribution including in the states of Arizona, Arkansas, California, Florida, Louisiana, Mississippi, North Carolina, North Dakota, Oklahoma, Pennsylvania, Georgia, Illinois, Kentucky, Tennessee, Texas, Washington, and Wisconsin.
- Location
- Hartland, WI
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11