PlainRecalls
FDA Devices Moderate Class II Ongoing

The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannula, depending on the model/outer diameter size of the forceps. Coagulation is achieved using electrosurgical energy under visualization. The device is to be used with bipolar outputs of compatible generators. The forceps' jaws are electrically isolated from each other enabling one jaw to act as a return electrode, eliminating the need for a return pad. To transect tissue, a cutting blade is actuated an

Reported: August 14, 2024 Initiated: July 1, 2024 #Z-2540-2024

Product Description

The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannula, depending on the model/outer diameter size of the forceps. Coagulation is achieved using electrosurgical energy under visualization. The device is to be used with bipolar outputs of compatible generators. The forceps' jaws are electrically isolated from each other enabling one jaw to act as a return electrode, eliminating the need for a return pad. To transect tissue, a cutting blade is actuated and moves between the electrode jaws. The connections (bipolar electrical signal) to the forceps are through either a hardwired cable (model 3006) or a connector at the bottom of the handle (models 3000 and 3005). The device includes mechanisms (switches and triggers) to provide coagulation using the bipolar electrical energy that reaches the forceps. The forceps are positioned in the desired location for grasping, coagulation, and transection of tissue. The anatomical structure to be coagulated is placed between the open forceps jaws, and once the Forceps Grip trigger is squeezed, the forceps' jaws are closed onto tissue, followed by coagulation. The device includes a blade that is mechanically advanced to perform tissue cutting as needed.

Reason for Recall

Fractures and breakages in packaging trays and Tyvek covers, which may result in a sterility breach

Details

Units Affected
100 unirs
Distribution
Worldwide distribution - US Nationwide and the countries of CA, DE, AU, SG, KR.
Location
Center Valley, PA

Frequently Asked Questions

What product was recalled?
The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannula, depending on the model/outer diameter size of the forceps. Coagulation is achieved using electrosurgical energy under visualization. The device is to be used with bipolar outputs of compatible generators. The forceps' jaws are electrically isolated from each other enabling one jaw to act as a return electrode, eliminating the need for a return pad. To transect tissue, a cutting blade is actuated and moves between the electrode jaws. The connections (bipolar electrical signal) to the forceps are through either a hardwired cable (model 3006) or a connector at the bottom of the handle (models 3000 and 3005). The device includes mechanisms (switches and triggers) to provide coagulation using the bipolar electrical energy that reaches the forceps. The forceps are positioned in the desired location for grasping, coagulation, and transection of tissue. The anatomical structure to be coagulated is placed between the open forceps jaws, and once the Forceps Grip trigger is squeezed, the forceps' jaws are closed onto tissue, followed by coagulation. The device includes a blade that is mechanically advanced to perform tissue cutting as needed.. Recalled by Olympus Corporation of the Americas. Units affected: 100 unirs.
Why was this product recalled?
Fractures and breakages in packaging trays and Tyvek covers, which may result in a sterility breach
Which agency issued this recall?
This recall was issued by the FDA Devices on August 14, 2024. Severity: Moderate. Recall number: Z-2540-2024.