CrossCHECK Non-Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-N3008; 3.0 x 10 mm, REF CCP-N3010; 3.0 x 12 mm, REF CCP-N3012; 3.0 x 14 mm, REF CCP-N3014l; 3.0 x 16 mm, REF CCP-N3016; 3.0 x 18 mm, REF CCP-N3018; 3.0 x 20 mm, REF CCP-N3020; 3.0 x 22 mm, REF CCP-N3022; 3.0 x 24 mm, REF CCP-N3024; 3.0 x 26 mm, REF CCP-N3026; 3.0 x 28 mm, REF CCP-N3028; 3.0 x 30 mm, REF CCP-N3030; 3.0 x 32 mm, REF CCP-N3032; 3.0 x 34 mm, REF CCP-N3034; 3.0 x 36 mm, REF CCP-N3036; 3.0 x 38 mm, REF CCP-N3038;

Reported: September 9, 2015 Initiated: July 21, 2015 #Z-2542-2015

Product Description

CrossCHECK Non-Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-N3008; 3.0 x 10 mm, REF CCP-N3010; 3.0 x 12 mm, REF CCP-N3012; 3.0 x 14 mm, REF CCP-N3014l; 3.0 x 16 mm, REF CCP-N3016; 3.0 x 18 mm, REF CCP-N3018; 3.0 x 20 mm, REF CCP-N3020; 3.0 x 22 mm, REF CCP-N3022; 3.0 x 24 mm, REF CCP-N3024; 3.0 x 26 mm, REF CCP-N3026; 3.0 x 28 mm, REF CCP-N3028; 3.0 x 30 mm, REF CCP-N3030; 3.0 x 32 mm, REF CCP-N3032; 3.0 x 34 mm, REF CCP-N3034; 3.0 x 36 mm, REF CCP-N3036; 3.0 x 38 mm, REF CCP-N3038; 3.0 x 40 mm, REF CCP-N3040; 3.5 x 8 mm, REF CCP-N3508; 3.5 x 10 mm, REF CCP-N3510; 3.5 x 12 mm, REF CCP-N3512; 3.5 14 mm, REF CCP-N3514; 3.5 x 16 mm, REF CCP-N3516; 3.5 x 18 mm, REF CCP-N3518; 3.5 x 20 mm, REF CCP-N3520; 3.5 x 22 mm, REF CCP-N3522; 3.5 x 24 mm, REF CCP-N3524; 3.5 x 26 mm, REF CCP-N3526; 3.5 x 28 mm, REF CCP-N3528; 3.5 x 30 mm, REF CCP-N3530; 3.5 x 32 mm, REF CCP-N3532; 3.5 x 34 mm, REF CCP-N3534; 3.5 x 36 mm, REF CCP-N3536; 3.5 x 38 mm, REF CCP-N3538; 3.5 x 40 mm, REF CCP-N3540. The CROSSCHECK Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet wrist and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.

Reason for Recall

These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.

Details

Units Affected
5687 units
Distribution
Nationwide Distribution
Location
Memphis, TN

Frequently Asked Questions

What product was recalled?
CrossCHECK Non-Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-N3008; 3.0 x 10 mm, REF CCP-N3010; 3.0 x 12 mm, REF CCP-N3012; 3.0 x 14 mm, REF CCP-N3014l; 3.0 x 16 mm, REF CCP-N3016; 3.0 x 18 mm, REF CCP-N3018; 3.0 x 20 mm, REF CCP-N3020; 3.0 x 22 mm, REF CCP-N3022; 3.0 x 24 mm, REF CCP-N3024; 3.0 x 26 mm, REF CCP-N3026; 3.0 x 28 mm, REF CCP-N3028; 3.0 x 30 mm, REF CCP-N3030; 3.0 x 32 mm, REF CCP-N3032; 3.0 x 34 mm, REF CCP-N3034; 3.0 x 36 mm, REF CCP-N3036; 3.0 x 38 mm, REF CCP-N3038; 3.0 x 40 mm, REF CCP-N3040; 3.5 x 8 mm, REF CCP-N3508; 3.5 x 10 mm, REF CCP-N3510; 3.5 x 12 mm, REF CCP-N3512; 3.5 14 mm, REF CCP-N3514; 3.5 x 16 mm, REF CCP-N3516; 3.5 x 18 mm, REF CCP-N3518; 3.5 x 20 mm, REF CCP-N3520; 3.5 x 22 mm, REF CCP-N3522; 3.5 x 24 mm, REF CCP-N3524; 3.5 x 26 mm, REF CCP-N3526; 3.5 x 28 mm, REF CCP-N3528; 3.5 x 30 mm, REF CCP-N3530; 3.5 x 32 mm, REF CCP-N3532; 3.5 x 34 mm, REF CCP-N3534; 3.5 x 36 mm, REF CCP-N3536; 3.5 x 38 mm, REF CCP-N3538; 3.5 x 40 mm, REF CCP-N3540. The CROSSCHECK Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet wrist and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.. Recalled by Wright Medical Technology, Inc.. Units affected: 5687 units.
Why was this product recalled?
These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 9, 2015. Severity: Moderate. Recall number: Z-2542-2015.