FDA Devices Moderate Class II Terminated

AD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kits: (a) REF DSG-90-2.3N (b) REF DSG-6.3-090-2.4N

Reported: September 25, 2019 Initiated: August 27, 2019 #Z-2544-2019

Product Description

AD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kits: (a) REF DSG-90-2.3N (b) REF DSG-6.3-090-2.4N

Reason for Recall

The inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.

Details

Units Affected
315 units
Distribution
NY, PA, WI, AZ, IL, CA, NH, GA, MA, MN, UT, MI, WA, FL, TX, OH, OR, PA, MD, and India, Korea, Taiwan, Canada, Hong Kong, Denmark, Russia, Singapore, Japan, Brazil. UK, Spain, Czech Republic, France
Location
Oak Creek, WI

Frequently Asked Questions

What product was recalled?
AD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kits: (a) REF DSG-90-2.3N (b) REF DSG-6.3-090-2.4N. Recalled by Ad-Tech Medical Instrument Corporation. Units affected: 315 units.
Why was this product recalled?
The inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 25, 2019. Severity: Moderate. Recall number: Z-2544-2019.