The product is a handheld ultraviolet-C germicidal wand and contains a narrow-diameter, tubular lamps with a peak emission wavelength in the 254 nm range.
Product Description
The product is a handheld ultraviolet-C germicidal wand and contains a narrow-diameter, tubular lamps with a peak emission wavelength in the 254 nm range.
Reason for Recall
The subject product emits unsafe level of UVC radiation (i.e., the product reaches the exposure limit for Risk Group 3 (high risk) in about 2.2 minutes). The product has a radiation safety defect, in that these radiation emissions are unnecessary to the accomplishment of the product s primary purpose of disinfecting objects and surfaces, and create a risk of skin and/or eye injuries to user and nearby persons.
Details
- Recalling Firm
- BigTree Sales Inc.
- Units Affected
- N/A
- Distribution
- U.S.A
- Location
- Arcadia, CA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11