HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kits or distributed separately: HeartMate 3 LVAS Implant Kit, US, Model: 106524US; HeartMate 3 LVAS Implant Kit, OUS, Model: 106524INT; HeartMate 3 System Controller, US, Model: 106531US; HeartMate 3 System Controller, OUS, Model: 106531INT; HeartMate 3 System Controller Low Flow 2.0, Global, Model: 106531LF2

Reported: August 21, 2024 Initiated: June 25, 2024 #Z-2583-2024

Product Description

HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kits or distributed separately: HeartMate 3 LVAS Implant Kit, US, Model: 106524US; HeartMate 3 LVAS Implant Kit, OUS, Model: 106524INT; HeartMate 3 System Controller, US, Model: 106531US; HeartMate 3 System Controller, OUS, Model: 106531INT; HeartMate 3 System Controller Low Flow 2.0, Global, Model: 106531LF2

Reason for Recall

Left ventricular assist system controller UI membrane/screen may lift along the edge of the controller housing, and if controller is exposed to fluid, ingress may occur, which may damage printed circuit board assemblies, which could result in unexpected and false positive alarms, loss of visual alarms, loss of user interface(use of buttons, loss of visual alarms,etc.), loss of power, and pump stop

Details

Recalling Firm
Thoratec LLC
Units Affected
694
Distribution
US: MN, PA, FL, IL, NY, DE, AR, NJ, NC, GA, WI, AZ, TN, MO, TX, MA, OH, VA, NE, CA, OK, CO, DC, IN, MI, CT, UT, IA, MD, OR, NM, ME, SC, KY, LA, WA, NV, HI, AL, MS, KS, WV. OUS: Algeria, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia & Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Korea, Kuwait, Lebanon, Lithuania, Macedonia, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan
Location
Pleasanton, CA

Frequently Asked Questions

What product was recalled?
HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kits or distributed separately: HeartMate 3 LVAS Implant Kit, US, Model: 106524US; HeartMate 3 LVAS Implant Kit, OUS, Model: 106524INT; HeartMate 3 System Controller, US, Model: 106531US; HeartMate 3 System Controller, OUS, Model: 106531INT; HeartMate 3 System Controller Low Flow 2.0, Global, Model: 106531LF2. Recalled by Thoratec LLC. Units affected: 694.
Why was this product recalled?
Left ventricular assist system controller UI membrane/screen may lift along the edge of the controller housing, and if controller is exposed to fluid, ingress may occur, which may damage printed circuit board assemblies, which could result in unexpected and false positive alarms, loss of visual alarms, loss of user interface(use of buttons, loss of visual alarms,etc.), loss of power, and pump stop
Which agency issued this recall?
This recall was issued by the FDA Devices on August 21, 2024. Severity: Moderate. Recall number: Z-2583-2024.