PlainRecalls
FDA Devices Moderate Class II Ongoing

t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)

Reported: October 1, 2025 Initiated: July 3, 2025 #Z-2621-2025

Product Description

t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)

Reason for Recall

an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.

Details

Units Affected
19
Distribution
International distribution to the country of Israel.
Location
San Diego, CA

Frequently Asked Questions

What product was recalled?
t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app). Recalled by Tandem Diabetes Care, Inc.. Units affected: 19.
Why was this product recalled?
an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 1, 2025. Severity: Moderate. Recall number: Z-2621-2025.