120 V Neptune 3 Rover, Model Number: 0703-001-000
Reported: August 31, 2016 Initiated: July 1, 2016 #Z-2630-2016
Product Description
120 V Neptune 3 Rover, Model Number: 0703-001-000
Reason for Recall
Stryker initiated a voluntary recall of the 120 V Neptune 3 Rover on 07/01/2016, due to a potential for the device to experience system errors causing the device to shutdown if high levels of electromagnetic interference are present in the operating room.
Details
- Recalling Firm
- Stryker Instruments Div. of Stryker Corporation
- Units Affected
- 78
- Distribution
- Domestic:MI, CA, ID, IA VA/DOD: None Foreign: None
- Location
- Portage, MI
Frequently Asked Questions
What product was recalled? ▼
120 V Neptune 3 Rover, Model Number: 0703-001-000. Recalled by Stryker Instruments Div. of Stryker Corporation. Units affected: 78.
Why was this product recalled? ▼
Stryker initiated a voluntary recall of the 120 V Neptune 3 Rover on 07/01/2016, due to a potential for the device to experience system errors causing the device to shutdown if high levels of electromagnetic interference are present in the operating room.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on August 31, 2016. Severity: Moderate. Recall number: Z-2630-2016.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11