PowerLoc MAX Power Injectable Infusion Set; Catalog Numbers 0131910 (UDI 00801741047329), 0131915 (UDI 00801741047336), 0131975 (UDI 00801741047343), 0132010 (UDI 00801741047350), 0132015 (UDI 00801741047367), 0132075 (UDI 00801741047374), 0132210 (UDI 00801741047381), 0132215 (UDI 00801741047398), 0132275 (UDI 00801741047404), 0141910 (UDI 00801741047411), 0141915 (UDI 00801741047428), 0141975 (UDI 00801741047435), 0142010 (UDI 00801741047442), 0142015 (UDI 00801741047459), 0142075 (UDI 00801
Product Description
PowerLoc MAX Power Injectable Infusion Set; Catalog Numbers 0131910 (UDI 00801741047329), 0131915 (UDI 00801741047336), 0131975 (UDI 00801741047343), 0132010 (UDI 00801741047350), 0132015 (UDI 00801741047367), 0132075 (UDI 00801741047374), 0132210 (UDI 00801741047381), 0132215 (UDI 00801741047398), 0132275 (UDI 00801741047404), 0141910 (UDI 00801741047411), 0141915 (UDI 00801741047428), 0141975 (UDI 00801741047435), 0142010 (UDI 00801741047442), 0142015 (UDI 00801741047459), 0142075 (UDI 00801741047466), 0142210 (UDI 00801741047473), 0142215 (UDI 00801741047480), 0142275 (UDI 00801741047497)
Reason for Recall
Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
Details
- Recalling Firm
- Becton Dickinson & Company
- Units Affected
- 5,296,132 total
- Distribution
- Worldwide Distribution
- Location
- Franklin Lakes, NJ
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11