Boston Scientific Extractor Pro RX-S Retrieval Balloon Catheter 12-15mm Below UPN: M00547340

Reported: July 12, 2017 Initiated: June 1, 2017 #Z-2674-2017

Product Description

Boston Scientific Extractor Pro RX-S Retrieval Balloon Catheter 12-15mm Below UPN: M00547340

Reason for Recall

Label on the affected devices incorrectly identifies the injection location as "above" the balloon when it should be below the balloon and "below" the balloon when it should be above the balloon

Details

Units Affected
32 units
Distribution
MI, MO Foreign: Austria France Great Britain Italy Spain Sweden
Location
Marlborough, MA

Frequently Asked Questions

What product was recalled?
Boston Scientific Extractor Pro RX-S Retrieval Balloon Catheter 12-15mm Below UPN: M00547340. Recalled by Boston Scientific Corporation. Units affected: 32 units.
Why was this product recalled?
Label on the affected devices incorrectly identifies the injection location as "above" the balloon when it should be below the balloon and "below" the balloon when it should be above the balloon
Which agency issued this recall?
This recall was issued by the FDA Devices on July 12, 2017. Severity: Moderate. Recall number: Z-2674-2017.