FDA Devices Moderate Class II Ongoing

Captiflex Snares, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snares flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loop.

Reported: August 12, 2020 Initiated: July 6, 2020 #Z-2764-2020

Product Description

Captiflex Snares, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snares flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loop.

Reason for Recall

BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.

Details

Distribution
World wide distribution.
Location
Marlborough, MA

Frequently Asked Questions

What product was recalled?
Captiflex Snares, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snares flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loop.. Recalled by Boston Scientific Corporation.
Why was this product recalled?
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 12, 2020. Severity: Moderate. Recall number: Z-2764-2020.