FDA Devices Moderate Class II Terminated

Non-Locking screws in the following lengths: (a) NEO(R) 0.27 - 10 mm, REF W27 ST010 (b) NEO(R) 0.27 - 12 mm, REF W27 ST012 (c) DRP 0.27 - 12 mm, REF W27 ST012 (d) NEO(R) 0.27 - 14 mm, REF W27 ST014 (e) NEO(R) 0.27 - 16 mm, REF W27 ST016 (f) NEO(R) 0.27 - 18 mm, REF W27 ST018 (f) DRP 0.27 - 18 mm, REF W27 ST018 The NEOVIEW(R) Plating System is intended for fixation of intra-articular and extra-articular fractures of the distal radius and reconstruction of the distal radius. It is composed

Reported: August 9, 2017 Initiated: July 13, 2017 #Z-2869-2017

Product Description

Non-Locking screws in the following lengths: (a) NEO(R) 0.27 - 10 mm, REF W27 ST010 (b) NEO(R) 0.27 - 12 mm, REF W27 ST012 (c) DRP 0.27 - 12 mm, REF W27 ST012 (d) NEO(R) 0.27 - 14 mm, REF W27 ST014 (e) NEO(R) 0.27 - 16 mm, REF W27 ST016 (f) NEO(R) 0.27 - 18 mm, REF W27 ST018 (f) DRP 0.27 - 18 mm, REF W27 ST018 The NEOVIEW(R) Plating System is intended for fixation of intra-articular and extra-articular fractures of the distal radius and reconstruction of the distal radius. It is composed of a NEOVIEW" plate in PEEK and NE0411 screws in Ti6Al4V that are available in Locking or Non-Locking versions. The NEOVIEW411 Screw Set consists of a standard set of NE0411 screws with different lengths suitable for the implantation of a NEOVIEW9 plate for an average anatomy

Reason for Recall

The firm received several customer complaints reporting some perioperative breakages of NE0411 Non-Locking screws when inserting the screw into the radius diaphysis

Details

Recalling Firm
In2Bones, SAS
Units Affected
287 screws
Distribution
Worldwide Diribution - US including Tennessee; Foreign - Italy, Portugal, Spain, Switzerland, UK, and France
Location
Ecully, N/A

Frequently Asked Questions

What product was recalled?
Non-Locking screws in the following lengths: (a) NEO(R) 0.27 - 10 mm, REF W27 ST010 (b) NEO(R) 0.27 - 12 mm, REF W27 ST012 (c) DRP 0.27 - 12 mm, REF W27 ST012 (d) NEO(R) 0.27 - 14 mm, REF W27 ST014 (e) NEO(R) 0.27 - 16 mm, REF W27 ST016 (f) NEO(R) 0.27 - 18 mm, REF W27 ST018 (f) DRP 0.27 - 18 mm, REF W27 ST018 The NEOVIEW(R) Plating System is intended for fixation of intra-articular and extra-articular fractures of the distal radius and reconstruction of the distal radius. It is composed of a NEOVIEW" plate in PEEK and NE0411 screws in Ti6Al4V that are available in Locking or Non-Locking versions. The NEOVIEW411 Screw Set consists of a standard set of NE0411 screws with different lengths suitable for the implantation of a NEOVIEW9 plate for an average anatomy. Recalled by In2Bones, SAS. Units affected: 287 screws.
Why was this product recalled?
The firm received several customer complaints reporting some perioperative breakages of NE0411 Non-Locking screws when inserting the screw into the radius diaphysis
Which agency issued this recall?
This recall was issued by the FDA Devices on August 9, 2017. Severity: Moderate. Recall number: Z-2869-2017.