Lactate Dehydrognease P-L, (UV), Catalog Number LD3818 - Product Usage: device intended for the quantitative in vitro determination of Lactate Dehydrogenase (LD) activity in serum and plasma. This product is suitable for use on the RX series instruments which includes the RX daytona and RX imola.
Product Description
Lactate Dehydrognease P-L, (UV), Catalog Number LD3818 - Product Usage: device intended for the quantitative in vitro determination of Lactate Dehydrogenase (LD) activity in serum and plasma. This product is suitable for use on the RX series instruments which includes the RX daytona and RX imola.
Reason for Recall
An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional pipette washes can be implemented as described in the technical bulletin. Interference to the Lactate Dehydrogenase (LOH) reagent would be observed as inconsistencies in Quality Control recovery, which may lead to a delay in running patient samples, or erroneous elevated test results.
Details
- Recalling Firm
- Randox Laboratories Ltd.
- Units Affected
- 4
- Distribution
- US Nationwide distribution including in the states of WV, and Puerto Rico.
- Location
- Crumlin (North), N/A
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11